An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. They run an SSH server accessible over the default port 22. The root account has a weak default password of ionadmin, and a password change policy for the root account is not enforced. Thus, an attacker with network connectivity can achieve root code execution. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.
CVE-2025-53963
CVSS base score
9.8 (v3.1)
Severity
CRITICAL
Attack vector
Network
Weakness
CWE-521
Last modified
November 4, 2026
CVSS vector
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Affected products
thermofisher ion torrent onetouch 2 firmware
References
- assets.thermofisher.com Product
- documents.thermofisher.com Vendor Advisory
- tools.thermofisher.cn Permissions Required